Basic Confocal Microscopy.pdf
(
364 KB
)
Pobierz
Basic Confocal Microscopy
UNIT 14.11
Confocal microscopy produces sharp im-
ages of structures within relatively thick speci-
mens (up to several hundred microns). It is
particularly useful for examining fluorescent
specimens. Thick fluorescent specimens
viewed with a conventional widefield fluores-
cent microscope appear blurry and lack contrast
because fluorophores throughout the entire
depth of the specimen are illuminated and fluo-
rescence signals are collected not only from the
plane of focus but also from areas above and
below. Confocal microscopes selectively col-
lect light from thin (
THE BASIS OF OPTICAL
SECTIONING
Confocal microscopes accomplish optical
sectioning by scanning the specimen with a
focused beam of light and collecting the fluo-
rescence signal from each spot via a spatial
filter (generally a pinhole aperture) that blocks
signals from out-of-focus areas of the speci-
men. The physical basis of optical sectioning
in fluorescence confocal microscopy is illus-
trated in Figure 14.11.2. A point light source
(typically a laser) evenly illuminates the back
focal plane of the objective, which focuses the
light to a diffraction-limited spot in the speci-
men. The irradiation is most intense at the focal
spot, although areas of the specimen above and
below the focal spot also are illuminated. Fluo-
rescent molecules excited by the incident light
emit fluorescence in all directions. The fluores-
cence collected by the objective comes to focus
in the image plane, which is conjugate (confo-
cal) with the focal plane in the specimen. A
pinhole aperture in the image plane allows
fluorescence from the illuminated spot in the
specimen to pass to the detector but blocks light
from out-of-focus areas.
The diameter of the pinhole determines how
much of the fluorescence emitted by the illu-
minated spot in the specimen is detected, and
the thickness of the optical section. From wave
optics we know that a point light source in the
plane of focus of an objective produces a three-
dimensional diffraction pattern in the image
plane. The cross section at the image plane is
an Airy disk (see Fig. 14.10.9), a circular dif-
fraction pattern with a bright central region.
The radius of the bright central region of the
Airy disk in the reference frame of the specimen
is given by
R
Airy
= 0.61
m) optical sections
representing single focal planes within the
specimen. Structures within the focal plane
appear more sharply defined than they would
with a conventional microscope because there
is essentially no flare of light from out-of-focus
areas. A three-dimensional view of the speci-
men can be reconstructed from a series of
optical sections at different depths.
The confocal microscope is the instrument
of choice for examining fluorescence-stained
cells in tissue slices or small, intact organisms
such as
Drosophila
(Fig. 14.11.1A,B) and ze-
brafish embryos. It is also useful for localizing
fluorescent-tagged molecules in dissociated
cells (Fig. 14.11.1C,D). Its sensitivity even al-
lows fluorescence in living specimens to be
monitored, making it feasible to follow the
movements in living cells of fluorescent probes
such as the green fluorescent protein (GFP; Fig.
14.11.1D). In addition, some types of confocal
microscopes can be configured to perform pho-
tobleach experiments (Fig. 14.11.1D) and to
photoactivate “caged” molecules (molecules
that are inactive until released with UV illumi-
nation).
Biologists use confocal microscopy in a
number of creative ways that are beyond the
scope of this article. The information presented
herein is intended to provide background and
practical tips needed to get started with confo-
cal microscopy. An excellent source of theoreti-
cal and technical information is the
Handbook
of Biological Confocal Microscopy
(1995; ed-
ited by J. Pawley). Also recommended are
Cell
Biological Applications of Confocal Micros-
copy
(1993; edited by B. Matsumoto), a good
source of practical information;
Confocal Mi-
croscopy
(1990; edited by T. Wilson), for theo-
retical background; and
Video Microscopy
(1997; Inoue and Spring) for fundamentals of
microscopy.
∼
1-
µ
λ
/NA
where
is the emission wavelength and NA is
the numerical aperture of the objective (see
UNIT
14.10
for a discussion of NA). At the image plane
(the location of the pinhole aperture), the radius
of the central region is
R
Airy
multiplied by the
magnification at that plane (for a more com-
plete explanation see Wilson, 1995).
Adjustment of the pinhole to a diameter
slightly less than the diameter of the central
region of the Airy disk allows most of the light
from the focal point to reach the detector and
reduces the background from out-of-focus ar-
eas by
λ
In situ
Hybridization
and Immuno-
histochemistry
∼
1000-fold relative to widefield micros-
Contributed by Carolyn L. Smith
Current Protocols in Molecular Biology
(1998) 14.11.1-14.11.12
Copyright © 1998 by John Wiley & Sons, Inc.
14.11.1
Supplement 44
Figure 14.11.1
Applications of laser scanning microscopy. (
A,B
) (From W. Oldenwald; see
Kamabadur et al., 1998.) 3-D analysis of thick specimens. Different neuronal populations of an
∼
250-
µ
m-thick Drosophila embryo were immunolabeled with antibodies against three transcription
factors. A,
, 0.8-NA objective using a detector pinhole
diameter of
∼
1.3 Airy units. Labeled neurons in the plane of focus appear sharply defined, while those
outside it are not visualized. B, projection (superimposition) of 65 optical sections collected at 2-
µ
m
intervals in the z axis. Neurons at different focal planes appear to overlap in this flattened image, but
are distinct in a 3-D reconstruction. (
C
) Localization of intracellular structures. Dissociated rat
fibroblasts were immunolabeled with anti-tubulin antibodies to visualize microtubules (green) and
stained with fluorescent probes for mitochondria (Mitotracker, red) and DNA (DAPI, blue). The image
is a projection of 20 optical sections collected at 0.3-
µ
m intervals in the z axis with 100
×
, 1.4-NA
objective. (
D
) Measuring molecular motility. In a living fibroblast expressing a Golgi membrane protein
(galactosyltransferase) fused to GFP (S65T-GFP), GFP fluorescence (green) localized to the Golgi
complex, shown superimposed on a DIC image of the cell. After the first image was collected, the
boxed region (yellow) was scanned with full laser power; this photobleached the GFP in the boxed
area as shown in the second image collected
2.5-
m optical section collected with 25
∼
µ
×
2 sec later. The rate of fluorescence recovery into the
photobleached zone (not illustrated) indicated that GFP-galactosyltransferase fusion is highly mobile
in Golgi membranes.
This black and white facsimile of the figure is intended only as a placeholder;
for full-color version of figure go to http://www.currentprotocols.com/colorfigures
∼
Basic Confocal
Microscopy
14.11.2
Supplement 44
Current Protocols in Molecular Biology
copy (Sandison et al., 1995). The separation of
the in-focus signal from the out-of-focus back-
ground achieved by a properly adjusted pinhole
is the principle advantage of confocal micros-
copy for examination of thick specimens (see
Fig. 14.11.1A,B).
Point illumination and the presence of a
pinhole in the detection light path also produces
improved lateral and axial resolution relative to
conventional microscopy (Table 14.11.1). The
actual extent of improvement depends on the
size of the pinhole. Near-maximal axial resolu-
tion is obtained with a pinhole radius
TYPES OF CONFOCAL
MICROSCOPES
Several types of confocal microscopes are
available, each having unique features and ad-
vantages. The types most commonly used for
examining fluorescence specimens are laser-
scanning confocal microscopes. These micro-
scopes, as their name implies, use lasers as light
sources and collect images by scanning the
laser beam across the specimen.
Lasers provide intense illumination within
a narrow range of wavelengths. The emission
wavelengths of several types of lasers, together
with the excitation spectra of familiar fluoro-
phores, are illustrated in Figure 14.11.3. Mixed
krypton-argon gas lasers are popular for multi-
wavelength confocal microscopy because they
emit at three well-separated wavelengths (488,
568, and 647 nm) that can be used to simulta-
neously image two or three fluorophores (e.g.,
FITC, lissamine rhodamine, and Cy5). The
disadvantage of krypton-argon lasers is that
×
R
Airy
whereas optimal lateral resolution is ob-
tained with a pinhole less than 0.3
∼
0.7
R
Airy
(Wilson, 1995). However, a pinhole smaller
than
×
R
Airy
significantly reduces the total
signal, a sacrifice that may not be worth the gain
in resolution, especially when imaging dim
samples. In fluorescence imaging, resolution
also is influenced by the emission and excita-
tion wavelengths (Table 14.11.1).
∼
0.7
×
photodetector
confocal pinhole
dichroic mirror
point
light
source
objective lens
specimen
focal plane
Figure 14.11.2
The basis of optical sectioning in confocal epifluorescence microscopy. Illumina-
tion from the point light source is reflected by the dichroic mirror and focused by the objective lens
to a diffraction-limited spot within the specimen. Fluorophores within the focal spot as well as in the
cone of light above and below it are excited, emitting fluorescence at a longer wavelength than the
incident light. The fluorescence captured by the objective passes through the dichroic mirror
because of its longer wavelength. The confocal pinhole allows fluorescence from the plane of focus
in the specimen to reach the photodetector but blocks fluorescence from areas above and below
the plane of focus. Redrawn from Shotton (1993).
In situ
Hybridization
and Immuno-
histochemistry
14.11.3
Current Protocols in Molecular Biology
Supplement 44
Table 14.11.1
Theoretical Resolutions of Confocal and Conventional Microscopes
a
Objective
λ
ex
/
λ
em
10
×
, 0.4 NA, air
40
×
, 0.85 NA, air
60
×
, 1.4 NA, oil
Lat. res.
Ax. res.
Lat. res.
Ax. res.
Lat. res.
Ax. res.
Confocal fluorescence microscope
488/518
0.55
4.50
0.26
0.99
0.16
0.56
568/590
0.64
5.17
0.30
1.09
0.18
0.64
647/677
0.72
5.88
0.34
1.28
0.21
0.72
Conventional fluorescence microscope
518
0.79
6.48
0.37
1.43
0.24
0.93
590
0.90
7.38
0.42
1.63
0.28
1.06
680
1.04
8.50
0.49
1.88
0.32
1.22
a
Data reprinted from Brelje et al. (1993) by permission of Academic Press.
λ
ex
and
λ
em
, excitation and emission
wavelengths; lat. res. and ax. res., lateral and axial resolutions.
Cascade blue
FITC TRITC LRSC
Cy5.18
100
80
60
40
20
0
300
350
400
450
500
550
600
650
700
Wavelength (nm)
argon
488 514
UV argon
351, 364
457
488 514 528
krypton-argon
488
568
647
krypton
478/482
520
568
647
676
helium-neon
543
595 633
helium-cadmium
325
354
422
Figure 14.11.3
Comparison of the emission wavelengths of various lasers and the excitation
spectra of representative fluorophores. The lasers most commonly used for laser-scanning confocal
microscopy are air-cooled argon (488 and 514 nm), krypton-argon, and helium-neon lasers. UV
argon lasers generally require water cooling and are more expensive. They may be configured to
provide only UV wavelengths (351 nm and 364 nm) or both UV and longer wavelengths. Data for
the excitation spectra of Cascade blue, fluorescein (FITC), tetramethylrhodamine (TRITC), lis-
samine rhodamine (LRSC), and cyanine 5.18 (Cy5.18) are from Wessendorf and Brelje (1993) and
were downloaded from the web page of Aryeh Weiss, http://optics.jct.ac.il/
∼
aryeh/Spectra. Modified
from Brelje et al. (1993).
Basic Confocal
Microscopy
14.11.4
Supplement 44
Current Protocols in Molecular Biology
their life spans are short (
2000 hr). Another
way to achieve multiwavelength excitation is
to combine the outputs of two or more lasers.
Several methods have been devised for scan-
ning the sample with the laser beam to illumi-
nate different positions in the specimen. The
most common method employs a pair of galva-
nometer mirrors to both scan the laser beam
across the specimen and collect the fluores-
cence emitted from the specimen (Fig.
14.11.4). One galvanometer mirror scans se-
quential spots along the
x
axis, and the second
mirror moves from line to line in the
y
axis. The
fluorescence emission is separated from the
illuminating beam by a dichroic beam splitter
and is directed to a photomultiplier tube which
collects the fluorescence produced as each spot
in the specimen is illuminated. The photodetec-
tor output is converted to a digital image that
can be displayed on a monitor and stored as a
digital image file for later analysis. Most laser-
scanning confocal microscopes have 8-bit digi-
∼
tizers that encode 256 gray levels, although
some recent models have 12- or 16-bit digitiz-
ers. Collection of a full-size image (typically
1024
2 sec. Laser-scan-
ning microscopes that employ galvanometer
mirror scanners sometimes are called “slow-
scan” microscopes because of their relatively
slow image acquisition rates. Slow-scan micro-
scopes are available from several sources (Bio-
Rad, Zeiss, Leica, Olympus, Nikon, Molecular
Dynamics, and Meridian; see
APPENDIX 4
).
The movements of the galvanometer mirrors
in laser-scanning microscopes are under the
control of a computer, providing flexibility in
the scanning pattern. For example, it is possible
to “zoom” a region of interest (visualize it at
higher magnification) by reducing the scan area
and the distances between sample points. In
addition, many laser-scanning microscopes
have the ability to repetitively scan a single line
or to “park” the scanner to monitor fluores-
cence at a single spot. The latter technique is
×
1024 pixels) takes
∼
krypton-argon
laser
variable
pinhole
emission
filter
specimen
line-selection filter
objective
photomultiplier
tube 2
neutral-density filter
X
mirror
Y
dichroic
beam
splitter 1
dichroic
beam
splitter 2
photomultiplier
tube 1
microscope
scanner
488-nm laser beam
lissamine rhodamine
FITC
568-nm laser beam
Figure 14.11.4
The light path of a laser-scanning confocal microscope set up for simultaneous
imaging of FITC and lissamine rhodamine. The 488-nm and 568-nm lines of a krypton-argon laser
are reflected by dichroic beam splitter 1 into the optical axis of the microscope. The scanner contains
two galvanometer mirrors, which generate the x and y axis movements of the beam. The beam is
reflected by a mirror into the objective which focuses the beam onto the specimen. The specimen
is scanned line by line in a raster pattern. Fluorescence emitted by the specimen as each spot is
illuminated travels the reverse path through the scanning system. The FITC fluorescence (peak at
520 nm) and lissamine rhodamine fluorescence (peak at 590 nm) pass through dichroic beam
splitter 1 to dichroic beam splitter 2, which transmits the lissamine rhodamine fluorescence to
photomultiplier tube 1 and reflects the FITC fluorescence to photomultiplier tube 2. A variable
pinhole in front of each photodetector blocks light from out-of-focus areas of the specimen while
allowing light from the illuminated spot to reach the detector.
In situ
Hybridization
and Immuno-
histochemistry
14.11.5
Current Protocols in Molecular Biology
Supplement 44
Plik z chomika:
vetman2010
Inne pliki z tego folderu:
Confocal Laser Scanning Microscopy.pdf
(968 KB)
A Confocal Laser Scanning Microscope investigation of different dental adhesives bonded to root canal dentine by K. Bitter, S. Paris, P. Martus, R. Schartner, A. M. Kielbassa.pdf
(244 KB)
Atlas of Confocal Laser Scanning In-vivo Microscopy in Opthalmology – Principles and Applications in Diagnostic and Therapeutic Ophtalmology by R.F.Guthoff, C. Baudouin, J. Stave.pdf
(17262 KB)
Basic Confocal Microscopy.pdf
(364 KB)
Confocal Microscopy Methods and Protocols by Stephen W. Paddock.pdf
(8147 KB)
Inne foldery tego chomika:
ARTYKUŁY
CHIRURGIA
Filmy
Galeria
GOSPODARSKIE
Zgłoś jeśli
naruszono regulamin